ETDC Jaipur

Measurement Capability

Sl. No.Parameter/measured quantity/Device Under CalibrationRangeCalibration and Measurement Capability (+)Standard equipment used
 Source
1.DC Voltage1mV -10mV0.1198% - 0.0140%

Multiproduct Calibrator,

Fluke-5520A

10mV -100mV0.0140% - 0.0036%
100mV -1V0.0036% - 0.0016%
1V – 1000V0.0016% - 0.0024%
2.DC Current10 uA  - 100 uA  0.2592% -0.0420%

Multiproduct Calibrator with Current Coil

Fluke-5520A

100 uA  - 1 mA0.0420% -0.0180%
1 mA –100 mA0.0180% -0.0156%
100mA -1A0.0156% - 0.0282%
1 A – 10A0.0282% - 0.0864%
10 A - 1000 A0.0864%  - 0.7085%
3.

AC Voltage

(45 Hz to 1kHz)

30 mV – 100mV0.1202% - 0.0265%

Multiproduct Calibrator,

Fluke-5520A

100mV to 100 V0.0265% - 0.0250%
100V - 1000 V0.0250% - 0.0362%
(1kHz – 100 kHz)1V - 30V0.0246% -0.1108%
4.

AC Current

(45 Hz to 1kHz)

30 uA - 100 uA0.5410% -0.2626%

Multiproduct Calibrator with Current Coil

Fluke-5520A

100 uA - 1 mA0.2626% -0.1356%
1 mA - 100 mA0.1356% -0.0841%
100 m A  - 10A0.0841% -0.1240%
50 Hz10 A - 1000 A0.1240% -0.71%
5.

AC Power

 

(1 phase at 50 Hz)

10V – 600V, 0.2A – 15A

 

UPF to 0.5 lag/ 0.5 lead

0.0645% - 0.3968%

Multiproduct Calibrator,

 

Fluke-5520A

6.DC Resistance1 mΩ - 10 mΩ0.9509% -0. 104%Standard Resistors L& N – 4222B, 4223B, YEW 2792, Decade Resistor Box
10 mΩ - 1 Ω0.104% -0. 0311%
1 Ω - 100 Ω0.0311% -0. 0059%
100 Ω - 100kΩ0.0059% - 0.015%
100k Ω - 1MΩ0.015% - 0.058%
1 MΩ - 100 GΩ0.058% - 1.542%
7.Capacitance (1kHz)100 PF – 1nF0.266% - 0.066%Standard Capacitors – GR 1409, Decade Capacitor Box
1nF – 100 nF0.066% - 0.077%
100 nF - 1 uF0.077% - 0.066%
8.Inductance (1kHz)100 uH – 1mH0.37% - 0.135%Standard Inductors – GR 1482, Decade Inductor Box
1mH - 10H0.135% - 0.137%
9.HF Voltage

10kHz–1GHz

 

-20dBm - +6dBm

(22.4mV  - 446mV)

1.26 dBmSignal Generator, Marconi-2022A
10.Frequency50Hz – 10 kHz28.87 ppm – 1.16 ppmSignal Generator, Marconi-2022A
10 kHz - 1GHz1.16 ppm
11.

Simulation for Calibration of Temp.Indicator/Controller/ Process Calibrator/

 

Thermocouple (J, K, T, R & S)

-10mV to 100mV

 

-200ºC to 1700ºC

0.0248ºC - 0.1331 ºC

Multiproduct Calibrator,

Fluke-5520A & ITS 90 by mV / ohm measurement

RTD (PT-100)

60 Ω to 400 Ω

 

-100ºC to 800ºC

0.008ºC - 0.042ºC

Measurement Capability (Measure)

Sl. No.Parameter/measured quantity/Device Under CalibrationRangeCalibration and Measurement Capability (+)Standard equipment used
 Measure   
1.DC Voltage1 mV –10 mV0.0631% - 0.0066%8 1/2 Digit Selfcal DMM, Wavetek-1271
10mV –100 mV0.0066% - 0.0019%
100 mV -100V0.0019% - 0.0012%
100V – 1000V0.0012% - 0.0015%
2.DC Current10 uA - 100 uA0.0770% -0.0136%8 1/2 Digit Selfcal DMM, Wavetek-1271, Fluke –87III & Clamp Meter
100 uA - 10 mA0.0136% -0.0074%
10mA - 1A0.0074% -0.0216%
1A - 10A0.0216% - 0.4705%
10A - 20A0.4705% - 3.40%
20 A - 1000 A3.40% - 2.32%
3.

AC Voltage

(45 Hz to 1kHz)

10 mV – 100mV0.1145% - 0.0377%8 1/2 Selfcal DMM, Wavetek-1271
100 mV – 100V0.0377% - 0.0291%
100V to 1000 V0.0291% - 0.0371%
(1kHz – 100 kHz)1V - 100V0.0291% - 0.1515%
4.

AC Current

(45 Hz to 1kHz)

30 uA - 100 mA0.1469% -0.0388%8 1/2 Selfcal DMM, Wavetek-1271, Fluke –87III & Clamp Meter
100 m A - 1A0.0388% -0.0876%
1 A - 10 A0.0876% - 1.3904%
10 A - 20 A1.3904% - 3.65%
20A –1000A3.65% -4.51%
5.DC Resistance1 mΩ - 100mΩ0.064% -0.0286%8 1/2 Selfcal DMM, Wavetek-1271 & Ultra High Resistance Meter, Advantest – R8340
100m Ω - 1 Ω0.0286% -0.0044%
1 Ω - 1k Ω0.0044% -0.0016%
1k Ω - 1MΩ0.0016% - 0.0020%
1 MΩ - 100 MΩ0.0020% - 0.0597%
100 MΩ - 100GΩ0.0597% - 2.500%
6.High Voltage (AC/DC)DCV: 1kV -5kV0.0015% - 1.0205%Electrostatic Voltmeter, ESH-28X
5kV -15kV1.0205% - 1.3760%
ACV: 1kV - 5kV0.1393% - 1.4046%
5kV -15kV1.4046% - 1.7372%
7.Capacitance (1KHz)100 PF – 1nF0.236% - 0.039%RLC Digibridge, GR-1693
1nF - 1 uF0.039% - 0.047%
8.Inductance (1KHz)100 uH – 1mH0.236% - 0.038%RLC Digibridge, GR-1693
1mH - 1H0.038% - 0.039%
1H - 10H0.039% - 0.067%
9.HF Voltage

10kHz–1GHz

 

-20dBm to +6dBm

(22.4mV - 446mV)

0.41dBm - 0.62dBmRF Millivolt meter URV-3
10.Frequency50 Hz – 1 kHz4.09 X 10-7– 5.18 X 10-8Electronic Counter, HP-5345A
1kHz – 100 MHz5.18 X 10-8 – 1.31 X 10-9
100 MHz – 500 MHz1.31 X 10-9 – 1.05 X 10-8
11.Time30Sec. – 60Min1 Sec.Electronic Counter, HP-5345A
12RTD (PT-100)60 Ω - 400 Ω 8 1/2 Selfcal DMM, Wavetek-1271 & & ITS 90 by mV / ohm measurement
-100ºC to 800ºC0.0063ºC - 0.0189ºC

Simulation for Calibration of Process Calibrator/ Thermocouple

(J, K, T, R, S, B,N & E)

-10mV to 100mV 
-200ºC to 100ºC0.0284ºC - 0.0142ºC
100ºC to 1700ºC0.0142ºC - 0.0687ºC

List of Major Facility

Facility / Equipment / ToolsMake and Model
Multiproduct Calibrator5520A, Fluke
Electronic Counter5345 A, HP
Signal Generator2022A, Marconi
RLC Digibridge1693, GR
Electrostatic VoltmeterESH28X, EIS
Selfcal DMM1271, Wavetek
RF MillivoltmeterURV-35, R&S
Megohm Bridge1644A, GR
Dry Block Temp. SourcePigasus 1200B
Slip Gauge SetM181/1,Micronix
Standard WeightsE2Class,Mettler
Electronic Weighing BalanceAT201,PG12001-S,Mettler
  

Major Products Calibrated / Project Done

Major Products Calibrated / Projects Done

Calibrators

DMM (5&1/2 digit)

Frequency counters

Signal Generators

Oscilloscope

LCR meter