Environmental Testing

Environment Test Lab.

ERTL(South) undertakes Environmental testing of systems & components as per MIL/ JSS/ IEC/IS/ Customer Specifications

Test Range/Accuracy/specifications

 

Dry Heat Test

 

Temp. : 40 to 150ºC/±2ºC, As per IS: 9000 Pt (III),

IEC 60068-2-2, JSS & MIL STDs

 

Cold Test

 

Temp.: -70ºC to +10º /±3ºC, IS:9000 Pt(II),

IEC 60068-2-1, JSS & MIL STDs

 

Damp Heat Test (Steady State)

 

Temp.: 40ºC ±2ºC, RH: 95±3%, IS: 9000 Pt (IV),

IEC 60068-2-78, JSS & MIL STDs

 

Damp Heat Test (Cyclic)

 

Temp.: 40ºC,55ºC,65ºC/±2ºC, RH: 95 ±3%, IS: 9000 Pt (V),

IEC 60068-2-30, JSS & MIL STDs

 

Corrosion Test

(Salt Spray)

 

Temp.: 35±2ºC, i) Seven Salt Solution ii) 20% or 5% NaCl

Solution, IS: 9000 Pt (XI), IEC 60068-2-11, JSS & MIL STDs

 

Temperature Cycling/ Change of Temperature

 

Temp.: - 65ºC to 180ºC , IS: 9000 Pt (XIV):1981,

IEC 60068-2-14

 

 

 

Vibration Test

(Sine & Random)

 

Max. Force. : 3000 Kgf (Peak Sine),3000 Kgf (RMS Random), Frequency : 5Hz-2KHz,

Displacement : 65.0 mm (P-P),

Acceleration : 1000 m/s2,

Sine : IS: 9000 Pt (VIII), IEC 60068-2-6, JSS STDs

Random : IEC 60068-2-35, 36,37&64

 

Fine leak Test

 

Fine Leak Test , Helium Tracer gas method as per

MIL 202,750 & 883

List of Major Facilities

Test Range/Accuracy/specifications

Equipment name and make

Specifications

Nitrogen Purged Burn-In system

Blue Star Tenney, India

Model:- T20-OVEN-(SP)

Temp: Ambient to 125 0C

Test Space: 80 cm x 106 cm x 76 cm

Thermal shock Test Chamber,

Weiss-Technik, Germany

Model :- TS 120/3

Temperature range :

Top chamber : +50 0C to +180 0C

Middle chamber : -10 0C to + 90 0C

Bottom chamber : - 80 0C to +70 0C.

Test Space : approx. 120 Litres

Programmable Humidity Chamber

Weiss-Technik, Germany

Model:- C-340/40

Temperature : -40 0C to 180 0C

RH: 10% to 98%

Test space: 335 Litres

Low/High Temperature Test Chamber

Weiss Technik, Germany

Model:- 305 SB/10 JU

Temp: Ambient to 85 0C

Test Space: 730 mm x 540 mm x 820 mm

Programmable Humidity Test Chamber

Weiss Technik, Germany

Model:-305 SB/10 JU

Temp: Ambient to 85 0C

Test space: 680 mm x 540 mm x 820 mm

Salt Spray Test Chamber

Votsch, Germany

Model :- VSC/KWT 450

Temp: Ambient to 55 0C

Test space: 750 mm x 850 mm x 570 mm

Climatic Test Chamber

Weins Technik, Germany

Model : WK 11 - 1000 / 40

Temp: -40 0C to 180 0C, RH: 10% to 98%

Test space:1000 mm x1000 mm x1000 mm

Low/High Temp Test Chamber

Votsch, Germany

Model:- VT-7100

Temp: -70 0C to 180 0C

Test space:1000 mm x1000 mm x1000 mm

Environmental Test Chamber

ACS,Italy

Model: Challenge 600

Temp: -10 0C to 850C

Rise rate: 4 0C /minute

 

Vibraton (3T) Test System

Saraswati Dynamics,India

Model:- SEV 300 VI

Frequency range : 5 Hz to 2500 Hz

Rated Force : 3000 Kgf peak sine &

3000 Kgf rms Random

Max. Acceleration: 100 g

Max. Displacement: 65 mm

Vibration Test System

Saraswati Dynamics, India

Model:- SEV 100

Frequency range : 5 Hz to 2000 Hz

Rated Force: 1000 Kgf (Peak Sine) &

720 Kgf (RMS Random)

Max. Acceleration: 70 g

Max. Displacement: 30 mm

Fine leak Detector

Agilent Technologies ,VS leak detector

Model:- VS MD 15

Minimum detectable Leak- Helium ,

atm-cc/sec - 5×10-12